Valve-in-valve transcatheter aortic valve implantation with fracturing of a failed small surgical aortic bioprosthesis: a case report
European Heart Journal - Case Reports

Abstract
Failure of a small surgical aortic bioprosthesis represents a challenging clinical scenario with valve-in-valve (ViV) transcatheter aortic valve implantation (TAVI) often resulting in patient-prosthesis mismatch. Bioprosthetic valve fracture (BVF) performed as a part of the ViV TAVI has recently emerged as an alternative approach with certain types of surgical bioprostheses.
An 81-year-old woman with a history of three surgical aortic valve procedures presented with heart failure. Aortic bioprosthesis degeneration with severe stenosis and moderate regurgitation was found. The patient was deemed a high-risk surgical candidate and the heart team decided that ViV TAVI was the preferred treatment option. Due to the very small 19 mm stented surgical aortic bioprosthesis Mitroflow 19 mm (Sorin Group, Italy) we decided to perform BVF as a part of ViV TAVI to prevent patient-prosthesis mismatch. Since this was the first BVF procedure in our centre, an
Bioprosthetic valve fracture together with ViV TAVI is a safe and effective emerging technique for treatment of small surgical aortic bioprosthesis failure. Bioprosthetic valve fracture allows marked oversizing of implanted self-expandable transcatheter aortic valves, leading to excellent haemodynamic and clinical results. An