Intermuscular technique for implantation of the subcutaneous implantable defibrillator: a propensity-matched case–control study
EP Europace Journal

Abstract
A previous randomized study demonstrated that the subcutaneous implantable cardioverter defibrillator (S-ICD) was noninferior to transvenous ICD with respect to device-related complications and inappropriate shocks. However, that was performed prior to the widespread adoption of pulse generator implantation in the intermuscular (IM) space instead of the traditional subcutaneous (SC) pocket. The aim of this analysis was to compare survival from device-related complications and inappropriate shocks between patients who underwent S-ICD implantation with the generator positioned in an IM position in comparison with an SC pocket.
We analysed 1577 consecutive patients who had undergone S-ICD implantation from 2013 to 2021 and were followed up until December 2021. Subcutaneous patients (
Our data showed the superiority of the IM S-ICD generator positioning in reducing device-related complications and inappropriate shocks.
Clinical Trial Registration: ClinicalTrials.gov; NCT02275637.