Predictors of venous obstruction following pacemaker or implantable cardioverter-defibrillator implantation: a contrast venographic study on 100 patients admitted for generator change, lead revision, or device upgrade
EP Europace Journal

Abstract
Venous obstruction following transvenous device implantation rarely cause immediate clinical problems. When lead revision or device upgrade is indicated, venous obstruction become a significant challenge. The aim of this study was to determine the predictors of venous obstruction after transvenous device implantation, and to asess likely effects of antiplatelet/anticoagulant drugs in preventing venous thrombosis.
Between March 2005 and July 2006, contrast venography was performed in 100 patients who were candidates for generator change, lead revision, or device upgrade. Vessel patency was graded as either completely obstructed, partially obstructed (>70%), or patent. The incidence of venous obstruction was 26%, with 9% of patients having total obstruction and 17% of patients exhibiting partial obstruction. No statistically significant differences between obstructed and non-obstructed patients were seen for age, sex, indication for device implantation, atrial fibrillation, cardiothoracic ratio, insulation material, operative technique, device type, and manufacturer (all
Total or partial obstruction of the access veins occurs relatively frequently after pacemaker or ICD implantation. Multiple pacing or ICD leads are associated with an increased risk of venous obstruction, whereas antiplatelet/anticoagulant therapy appears to have a preventive effect on development of access vein thrombosis.
Contributors

Mohammad Hossein Nikoo
Author

Amir Farjam Fazelifar
Author

Abolfath Alizadeh
Author

Zahra Emkanjoo
Author

Mohammad Ali Sadr-Ameli
Author