Performance of handheld electrocardiogram devices to detect atrial fibrillation in a cardiology and geriatric ward setting
EP Europace Journal

Abstract
To determine the usability, accuracy, and cost-effectiveness of two handheld single-lead electrocardiogram (ECG) devices for atrial fibrillation (AF) screening in a hospital population with an increased risk for AF.
Hospitalized patients (
Using AliveCor or MyDiagnostick handheld recorders requires a structured screening strategy to be effective and cost-effective in a hospital setting. It must exclude patients with implanted devices and known AF, and requires targeted additional 12-lead ECGs to optimize specificity. Under these circumstances, the expenses per diagnosed new AF patient and preventable stroke are reasonable.
Contributors

Zina Raymaekers
Author

Mark Lutin
Author

Johan Vijgen
Author

Dagmara Dilling-Boer
Author

Pieter Koopman
Author

Joris Schurmans
Author

Philippe Vanduynhoven
Author

Paul Dendale
Author

Hein Heidbuchel
Author
You may be interested in