Long-term outcome of transvenous pacemaker implantation in infants: a retrospective cohort study
EP Europace Journal

Abstract
Evaluation of long-term outcome of transvenous pacemaker (PM) implantation in infants.
A retrospective analysis of all transvenous PM implantations in infants <10 kg between September 1997 and October 2001 was made. Indications for PM implantation, age at implantation, and determinants of long-term outcome including cardiac function, PM function, and PM (system) complications were noted. Seven patients underwent transvenous VVI(R) PM implantation. Median age at implantation was 3 days (range: 1 day to 14 months), median weight 3.5 kg (range: 2.3–8.7 kg), and median follow-up 14 years (range: 12.3–16.3 years). Pacemaker indications were congenital complete atrioventricular block (
Transvenous PM implantation in infants (<10 kg) is associated with a high incidence of vascular occlusion, thrombosis, and severe atrioventricular valve regurgitation during long-term follow-up. We advocate an epicardial approach for PM implantation in small children.
Contributors

Laura M. Vos
Author

Janneke A.E. Kammeraad
Author

Matthias W. Freund
Author

Andreas C. Blank
Author

Johannes M.P.J. Breur
Author
You may be interested in