Axillary vein access for pacemaker implantation is uncommon in many centres because of the lack of training in this technique. We assessed whether the introduction of the axillary vein technique was safe and efficient as compared with cephalic vein access, in a centre where no operators had any previous experience in axillary vein puncture.
Patients undergoing pacemaker implantation were randomized to axillary or cephalic vein access. All three operators had no experience nor training in axillary vein puncture, and self-learned the technique by reading a published review. Axillary vein puncture was fluoroscopy-guided without contrast venography. Cephalic access was performed by dissection of delto-pectoral groove. Venous access success, venous access duration (from skin incision to guidewire or lead in superior vena cava), procedure duration, X-ray exposure, and peri-procedural (1 month) complications were recorded. results We randomized 74 consecutive patients to axillary (
Self-taught axillary vein puncture for pacemaker implantation seems immediately safe and faster than cephalic vein access, when performed by electrophysiologists trained to pacemaker implantation but not to axillary vein puncture.