A novel assessment of local impedance during catheter ablation: initial experience in humans comparing local and generator measurements
EP Europace Journal

Abstract
A novel measure of local impedance (LI) has been found to predict lesion formation during radiofrequency current (RFC) catheter ablation. The aim of this study was to investigate the utility of this novel approach, while comparing LI to the well-established generator impedance (GI).
In 25 consecutive patients with a history of atrial fibrillation, catheter ablation was guided by a 3D-mapping system measuring LI in addition to GI via an ablation catheter tip with three incorporated mini-electrodes. Local impedance and GI before and during RFC applications were studied. In total, 381 RFC applications were analysed. The baseline LI was higher in high-voltage areas (>0.5 mV; LI: 110.5 ± 13.7 Ω) when compared with intermediate-voltage sites (0.1–0.5 mV; 90.9 ± 10.1 Ω,
Local impedance can be monitored during ablation. Compared with the sole use of GI, baseline LI is a better predictor of impedance drops during ablation and may provide useful insights regarding lesion formation. However, further studies are needed to investigate if this novel approach is useful to guide catheter ablation.