Lead extraction for reduction of chronic pain related to cardiovascular implantable electronic device
EP Europace Journal

Abstract
Chronic pain at the cardiovascular implantable electronic device (CIED) generator or lead insertion site that is not otherwise manageable carries a IIA indication for extraction. However, limited data exist evaluating causes of pain and outcomes of extraction in eliminating pain. A multi-centre retrospective observational study was conducted to evaluate outcomes of patients undergoing device extraction for treatment of chronic device pain.
Twenty-seven out of 2188 lead extraction candidates (1.3%) met the chronic pain IIA indication for extraction [50 ± 16 years; 14 (51%) women]. Onset, severity, triggers, and pain management were measured before and after extraction. Device type, procedure done (with/without reimplantation), and positive tissue cultures were noted. Pain was reported as constant (
Chronic pain at the CIED generator site can present as chronic or movement-triggered pain, and can be due to subclinical infection or a poorly formed device pocket. Extraction relieved constant and intermittent pain in two-thirds of patients. Extraction appears less successful in eliminating pain in women who undergo subsequent reimplantation.
Contributors

Kathryn Lewis
Author

Mingyan Dai
Author

Kristen K Patton
Author

Yong-Mei Cha
Author

Travis Pollema
Author

Gregory K Feld
Author

Ulrika Birgersdotter-Green
Author
University of California San Diego La Jolla , United States of America

Victor Pretorius
Author
You may be interested in