Electrocardiographic characterization of non-selective His-bundle pacing: validation of novel diagnostic criteria
EP Europace Journal

Abstract
Permanent His-bundle (HB) pacing is usually accompanied by simultaneous capture of the adjacent right ventricular (RV) myocardium—this is described as a non-selective (ns)-HB pacing. It is of clinical importance to confirm HB capture using standard electrocardiogram (ECG). Our aim was to identify ECG criteria for loss of HB capture during ns-HB pacing.
Patients with permanent HB pacing were recruited. Electrocardiograms during ns-HB pacing and loss of HB capture (RV-only capture) were obtained. Electrocardiogram criteria for loss/presence of HB capture were identified. In the validation phase, these criteria and the ‘HB ECG algorithm’ were tested using a separate, sizable set of ECGs. A total of 353 ECG (226 ns-HB and 128 RV-only) were obtained from 226 patients with permanent HB pacing devices. QRS notch/slur in left ventricular leads and R-wave peak time (RWPT) in lead V6 were identified as the best features for differentiation. The ‘HB ECG algorithm’ based on these features correctly classified 87.1% of cases with sensitivity and specificity of 93.2% and 83.9%, respectively. The criteria for definitive diagnosis of ns-HB capture (no QRS slur/notch in Leads I, V1, V4–V6, and the V6 RWPT ≤ 100 ms) presented 100% specificity.
A novel ECG algorithm for the diagnosis of loss of HB capture and criteria for definitive confirmation of HB capture were formulated and validated. The algorithm might be useful during follow-up and the criteria for definitive confirmation of ns-HB capture offer a simple and reliable ancillary procedural endpoint during HB device implantation.
Contributors

Paweł Moskal
Author

Karol Curila
Author

Kamil Fijorek
Author

Piotr Kukla
Author

Agnieszka Bednarek
Author

Grzegorz Kiełbasa
Author

Adam Bednarski
Author

Adrian Baranchuk
Author

Danuta Czarnecka
Author