Accuracy of fully automated right ventricular quantification software with 3D echocardiography: direct comparison with cardiac magnetic resonance and semi-automated quantification software
European Heart Journal - Cardiovascular Imaging

Abstract
The aim of this study was to determine the accuracy and reproducibility of a novel, fully automated 3D echocardiography (3DE) right ventricular (RV) quantification software compared with cardiac magnetic resonance (CMR) and semi-automated 3DE RV quantification software.
RV volumes and the RV ejection fraction (RVEF) were measured using a fully automated software (Philips), a semi-automated software (TomTec), and CMR in 100 patients who had undergone both CMR and 3DE examinations on the same day. The feasibility of the fully automated software was 91%. Although the fully automated software, without any manual editing, significantly underestimated RV end-diastolic volume (bias: −12.6 mL,
A novel, fully automated 3DE RV quantification software underestimated RV volumes but successfully approximated RVEF when compared with CMR. No inferiority of this software was observed when compared with the semi-automated software. Rapid analysis and higher reproducibility also support the routine adoption of this method in the daily clinical workflow.
Contributors