Construct validity of automated assessment of invasively measured hemodynamics during transcatheter aortic valve replacement

European Heart Journal - Digital Health

20 June 2025
Organised by: Logo
ESC Journals Interventional Cardiology VALVULAR, MYOCARDIAL, PERICARDIAL, PULMONARY, CONGENITAL HEART DISEASE Valvular Heart Disease

Abstract

AbstractAims

Paravalvular regurgitation (PVR) is frequently observed following Transcatheter Aortic Valve Replacement (TAVR). Periprocedural monitoring of invasive hemodynamics has shown promise for diagnosis of PVR, but automated software options are lacking. We aimed to develop a rule-based algorithm for automated assessment of hemodynamic indices of PVR, and evaluate its construct validity and discriminatory value for cardiac magnetic resonance (CMR)-derived relevant PVR compared to standard manual hemodynamic assessment.

Methods and results

Left ventricular and aortic pressures were invasively measured during TAVR using fluid-filled pigtail catheters. To evaluate construct validity of automated vs. manual assessment of invasive hemodynamics, we compared (i) proportion of cardiac cycles affected by arrhythmias/noise, (ii) pressure gradients, and (iii) PVR indices. Additionally, we compared the discriminatory value of automatically and manually determined PVR indices for CMR-determined relevant PVR at 30-days. In total, 77 patients were enrolled (664 cardiac cycles). Automated filtering of cardiac cycles affected by arrhythmias/noise had a high sensitivity (95.2%) and specificity (86.4%). In addition, excellent agreement was observed between automated and manual computation of mean gradients pre- and post-TAVR [39.3 ± 12.1 vs. 37.5 ± 11.9 mmHg, intra-class correlation coefficient (ICC): 0.916; 1.92 ± 5.87 vs. 1.14 ± 5.89, ICC: 0.957, respectively], and PVR indices [diastolic delta (DD): 41.7 ± 12.4 vs. 40.6 ± 12.3 mmHg, ICC: 0.982, respectively]. Automated and manual assessment of DD showed comparable discriminatory value for relevant PVR [area under the curve (AUC): 0.81 vs. 0.80, respectively].

Conclusion

Rule-based, automated assessment of hemodynamic indices of PVR showed excellent construct validity and discriminatory value for CMR-determined relevant PVR, supporting its use for real-time evaluation and risk stratification in TAVR patients.

Contributors

Niels A Stens
Niels A Stens

Author

Radboud University Medical Centre Nijmegen , Netherlands (The)

Maxim J P Rooijakkers
Maxim J P Rooijakkers

Author

Radboud University Medical Centre Nijmegen , Netherlands (The)

Niels van Royen
Niels van Royen

Author

Radboud University Nijmegen Nijmegen , Netherlands (The)