Impact of wall thickness on the tissue cooling effect of cryoballoon ablation
EP Europace Journal

Abstract
Understanding of the tissue cooling properties of cryoballoon ablation during pulmonary vein (PV) isolation is lacking. The purpose of this study was to delineate the depth of the tissue cooling effect during cryoballoon freezing at the pulmonary venous ostium.
A left atrial-PV model was constructed using a three-dimensional printer with data from a patient to which porcine thigh muscle of various thicknesses could be affixed. The model was placed in a 37°C water tank with a PV water flow at a rate that mimicked biological blood flow. Cryofreezing at the PV ostium was performed five times each for sliced porcine thigh muscle of 2, 4, and 6 mm thickness, and sliced muscle cooling on the side opposite the balloon was monitored. The cooling effect was assessed using the average temperature of 12 evenly distributed thermocouples covering the roof region of the left superior PV. Tissue cooling effects were in the order of the 2, 4, and 6 mm thicknesses, with an average temperature of −41.4 ± 4.2°C for 2 mm, −33.0 ± 4.0°C for 4 mm, and 8.0 ± 8.7°C for 6 mm at 180 s (
The cooling effect of cryoballoon freezing is weaker in the deeper layers. Cryoballoon ablation should be performed with consideration to myocardial thickness.
Contributors

Masaharu Masuda
Author

Yasuhiro Matsuda
Author

Hiroyuki Uematsu
Author

Mizuki Nishijima
Author

Shin Okamoto
Author

Takayuki Ishihara
Author

Kiyonori Nanto
Author

Takuya Tsujimura
Author

Yosuke Hata
Author

Toshiaki Mano
Author