Substernal lead removal: experience from the EV ICD pivotal study

EP Europace Journal

24 May 2024
Organised by: Logo
ESC Journals

Abstract

AbstractBackground

The extravascular implantable cardioverter defibrillator (EV ICD) has recently emerged as a new option for guideline indicated ICD patients. Previous study results from the EV ICD Pilot and Pivotal studies have demonstrated that the device can provide safe and effective therapy through chronic follow up. However, less has been reported regarding EV ICD lead removal.

Purpose

To report on the experience of lead removal in the EV ICD Pivotal Study.

Methods

The EV ICD Pivotal study was an international, prospective, single-arm, premarket clinical study. Patients with a class I or IIa indication for a single-chamber ICD per ESC or ACC/AHA/HRS guidelines were enrolled. Lead removal summary data was reported.

Results

Of 316 patients with an implant attempt, (74.7% male, age 53.8±13.1 years, 82% primary prevention, LVEF of 38.9%±15.4%, and NYHA Class I [23.7%] or II/III [65.5%]) 299 were fully implanted; 17 (5.7%) of which had a lead removal procedure (14–752 days post-implant, at 13 sites) with 3 of them having been implanted for more than 1 year. Reasons for lead removal were lead dislodgement (n=7), infection (n=5), high impedance alert due to lead fracture (n=2), pacing indication (n=1), cardiac transplant (n=1), or scheduled sternotomy (n=1). New EV ICD leads were placed in 7/17 (41.2%) patients. No serious adverse events were adjudicated as related to the explant procedures. There was a single observation of incision site haemorrhage, that was resolved with wound care. Additional details on lead removal will be included at the time of presentation.

Conclusions

The EV ICD lead was successfully removed 14-752 days post implant. Nine patients were re-implanted with either extravascular or transvenous systems. Longer term extraction data collection is ongoing through the EV ICD Pivotal study and through the EV ICD post-approval study and will continue to provide additional insights.

Lead removed 24 months post-implant.

Contributors

A Amin
A Amin

Author

OhioHealth Riverside Methodist Hospital Columbus , United States of America

I Crozier
I Crozier

Author

Christchurch Hospital Christchurch , New Zealand

C Leclercq
C Leclercq

Author

Hospital Pontchaillou of Rennes Rennes , France

J Y S Chan
J Y S Chan

Author

Prince of Wales Hospital Hong Kong, Chinese University of Hong Kong Sha Tin , Hong Kong

H Roukoz
H Roukoz

Author

M Biffi
M Biffi

Author

S. Orsola-Malpighi Policlinic Bologna , Italy

V Sagi
V Sagi

Author

C Wiggenhorn
C Wiggenhorn

Author

Medtronic Mounds View , United States of America

P Friedman
P Friedman

Author

Mayo Clinic Rochester , United States of America