Predicting worsening heart failure hospitalizations in patients with implantable cardioverter defibrillators: is it all about alerts? A pooled analysis of nine trials
EP Europace Journal

Abstract
To predict worsening heart failure hospitalizations (WHFHs) in patients with implantable defibrillators and remote monitoring, the HeartInsight algorithm (Biotronik, Berlin, Germany) calculates a heart failure (HF) score combining seven physiologic parameters: 24 h heart rate (HR), nocturnal HR, HR variability, atrial tachyarrhythmia, ventricular extrasystoles, patient activity, and thoracic impedance. We compared temporal trends of the HF score and its components 12 weeks before a WHFH with 12-week trends in patients without WHFH, to assess whether trends indicate deteriorating HF regardless of alert status.
Data from nine clinical trials were pooled, including 2050 patients with a defibrillator capable of atrial sensing, ejection fraction ≤ 35%, NYHA class II/III, no long-standing atrial fibrillation, and 369 WHFH from 259 patients. The mean HF score was higher in the WHFH group than in the no WHFH group (42.3 ± 26.1 vs. 30.7 ± 20.6,
The HF score was significantly higher at, and further increased during 12 weeks before WHFH, as compared to the no WHFH group, with seven components showing different behaviour and contribution. Temporal trends of HF score may serve as a quantitative estimate of HF condition and evolution prior to WHFH.
Contributors

Alan Bulava
Author

Alessio Gargaro
Author

Tobias Timmel
Author

Daniele Giacopelli
Author

Laurence Guédon-Moreau
Author
You may be interested in