Procedural outcome and follow-up of stylet-driven leads compared with lumenless leads for left bundle branch area pacing
EP Europace Journal

Abstract
Left bundle branch area pacing (LBBAP) is most often delivered using lumenless leads (LLLs), but may also be performed using stylet-driven leads (SDLs). There are limited reports on the comparison of these tools, mainly limited to reports describing initial operator experience or without detailed procedural data. Our aim was to perform an in-depth comparison of SDLs and LLLs for LBBAP at implantation and follow-up in a larger cohort of patients with experience that extends beyond that of the initial learning curve.
A total of 306 consecutive patients (age 77 ± 11 years, 183 males) undergoing LBBAP implantation at a single centre were prospectively included. The population was split into two groups of 153 patients based on the initial use of an SDL (from 4 manufacturers) or an LLL. After having discounted the initial learning curve of 50 patients, there was no difference in the success rate between the initial use of lead type (96.0% with SDL vs. 94.3% with LLL,
Left bundle branch area pacing may be safely and effectively performed using either LLLs or SDLs, which provides implanters with alternatives for delivering this therapy.