Cardiac resynchronization therapy non-responder to responder conversion rate in the MORE-CRT MPP trial
EP Europace Journal

Abstract
To assess the impact of MultiPoint™ Pacing (MPP) in cardiac resynchronization therapy (CRT) non-responders after 6 months of standard biventricular pacing (BiVP).
The trial enrolled 5850 patients who planned to receive a CRT device. The echocardiography core laboratory assessed CRT response before implant and after 6 months of BiVP; non-response to BiVP was defined as <15% relative reduction in left ventricular end-systolic volume (LVESV). Echocardiographic non-responders were randomized in a 1:1 ratio to receive MPP (541 patients) or continued BiVP (570 patients) for an additional 6 months and evaluated the conversion rate to the echocardiographic response. The characteristics of both groups at randomization were comparable. The percentage of non-responder patients who became responders to CRT therapy was 29.4% in the MPP arm and 30.4% in the BIVP arm (
Our trial shows that ∼30% of patients, who do not respond to CRT in the first 6 months, experience significant reverse remodelling in the following 6 months. This finding suggests that CRT benefit may be delayed or slowly incremental in a relevant proportion of patients and that the percentage of CRT responders may be higher than what has been described in short-/middle-term studies. MultiPoint™ Pacing does not improve CRT response in non-responders to BiVP, even with MPP-AS.
Contributors

Peter Paul Delnoy
Author

Christopher A Rinaldi
Author

Johannes Sperzel
Author

Leonardo Calò
Author

Joaquin Fernandez Concha
Author

Antonio Fusco
Author

Faisal Al Samadi
Author

Kwangdeok Lee
Author

Bernard Thibault
Author
You may be interested in




