Smartwatch-derived heart rate variability: a head-to-head comparison with the gold standard in cardiovascular disease
European Heart Journal - Digital Health

Abstract
We aimed to investigate the concordance between heart rate variability (HRV) derived from the photoplethysmographic (PPG) signal of a commercially available smartwatch compared with the gold-standard high-resolution electrocardiogram (ECG)-derived HRV in patients with cardiovascular disease.
We prospectively enrolled 104 survivors of acute ST-elevation myocardial infarction, 129 patients after an ischaemic stroke, and 30 controls. All subjects underwent simultaneous recording of a smartwatch (Garmin vivoactive 4; Garmin Ltd, Olathe, KS, USA)-derived PPG signal and a high-resolution (1000 Hz) ECG for 30 min under standardized conditions. HRV measures in time and frequency domain, non-linear measures, as well as deceleration capacity (DC) were calculated according to previously published technologies from both signals. Lin’s concordance correlation coefficient (
Smartwatch-derived HRV provides a practical alternative with excellent accuracy compared with ECG-based HRV for global markers and those characterizing lower frequency components. However, caution is warranted with HRV markers that predominantly assess short-term variability.
Contributors

Fabian Theurl
Author

Michael Schreinlechner
Author

Nikolay Sappler
Author

Michael Toifl
Author

Theresa Dolejsi
Author

Florian Hofer
Author

Celine Massmann
Author

Christian Steinbring
Author

Silvia Komarek
Author

Kurt Mölgg
Author

Benjamin Dejakum
Author

Christian Böhme
Author

Rudolf Kirchmair
Author

Sebastian Reinstadler
Author
You may be interested in