Outcome and safety of intraoperative defibrillation testing during device replacement: the Simpler trial
EP Europace Journal

Abstract
Intraoperative defibrillation testing (DT) during implant or replacement of implantable cardioverter-defibrillators (ICDs) has been a matter of debate for many years. This debate was put to rest by the Simple and Nordic ICD trials, and the practice of testing during new implantations has essentially been almost abandoned. Old registries demonstrated an increased incidence of significant findings in DT during replacements. The aim of the present study was to evaluate frequency of significant findings and safety of DT in subjects undergoing device replacement.
A prospective observational multi-centre study included consecutive patients undergoing ICD generator replacement. The primary outcome was a failure to terminate induced ventricular fibrillation (VF) with a single shock 10 J below the maximal capacity of the device. Secondary outcomes included complications of DT. Patients were followed-up at 1- and 6-months post-procedure. A total of 92 patients were eligible, and consented to the study, of which 84 underwent DT during battery replacement. The median age was 68 years and 79.8% were males. Induction of VF was successful in 84 patients as was a successful conversion on the first attempt in all. There were no procedure-related complications. During follow up one patient had two appropriate ICD shock events. In four patients, ICD programming was changed. None suffered inappropriate shock. There was no evidence of lead malfunction. Two deaths occurred, none of which was related to arrhythmia.
The present study found DT was not associated with complications in patients undergoing ICD generator replacement but produced no clinically important information.
Contributors

Eyal Nof
Author

Moshe Rav Acha
Author

Roy Beinart
Author

Valentina Kutyifa
Author

Béla Merkely
Author

Ehud Regev
Author

Yong-Mei Cha
Author

Tal Ovdat
Author

Robert Klempfner
Author

Michael Glikson
Author
You may be interested in