Outcomes after cryoballoon ablation of paroxysmal atrial fibrillation with the PolarX or the Arctic Front Advance Pro: a prospective multicentre experience
EP Europace Journal

Abstract
The aim of this study was to compare procedural efficacy and safety, including 1-year freedom from AF recurrence, between the novel cryoballoon system PolarX (Boston Scientific) and the Arctic Front Advance Pro (AFA-Pro) (Medtronic), in patients with paroxysmal AF undergoing PVI.
This multicentre prospective observational study included 267 consecutive patients undergoing a first cryoablation procedure for paroxysmal AF (137 PolarX, 130 AFA-Pro). Kaplan-Meier curves with the log-rank test was used to compare the 1-year freedom from AF recurrence between both groups. Multivariate Cox model was performed to evaluate whether the type of procedure (PolarX vs. AFA-Pro) had an impact on the occurrence of AF recurrences after adjustment on potentially confounding factors. The PolarX reaches lower temperatures than the AFA-Pro (LSPV 52 ± 5, vs. 59 ± 6; LIPV 49 ± 6 vs. 56 ± 6; right superior pulmonary vein: 49 ± 6 vs. 57 ± 7; right inferior pulmonary vein: 52 ± 6 vs. 59 ± 6;
PolarX and AFA-Pro have comparable efficacy and safety profiles for pulmonary veins isolation in paroxysmal atrial fibrillation.
Contributors

Nikita Tanese
Author

Pascal Defaye
Author

Sophie Jacob
Author

Mouna Ben Kilani
Author

Gianbattista Chierchia
Author

Sandrine Venier
Author

Christèle Cardin
Author

Peggy Jacon
Author

Stéphane Combes
Author

Elodie Deschamps
Author

Carlo de Asmundis
Author
You may be interested in