Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation
EP Europace Journal

Abstract
Although mobile health tools using photoplethysmography (PPG) technology have been validated for the detection of atrial fibrillation (AF), their utility for heart rate assessment during AF remains unclear. Therefore, we aimed to evaluate the accuracy of continuous PPG-based 1 min mean heart rate assessment during AF.
Persistent AF patients were provided with Holter electrocardiography (ECG) (for ≥24 h) simultaneously with a PPG-equipped smartwatch. Both the PPG-based smartwatch and Holter ECG automatically and continuously monitored patients’ heart rate/rhythm. ECG and PPG recordings were synchronized and divided into 1 min segments, from which a PPG-based and an ECG-based average heart rate estimation were extracted. In total, 47 661 simultaneous ECG and PPG 1 min heart rate segments were analysed in 50 patients (34% women, age 73 ± 8 years). The agreement between ECG-determined and PPG-determined 1 min mean heart rate was high [root mean squared error (RMSE): 4.7 bpm]. The 1 min mean heart rate estimated using PPG was accurate within ±10% in 93.7% of the corresponding ECG-derived 1 min mean heart rate segments. PPG-based 1 min mean heart rate estimation was more often accurate during night-time (97%) than day-time (91%,
During persistent AF, continuous PPG-based 1 min mean heart rate assessment is feasible in 60% of the analysed period and shows high accuracy compared with Holter ECG for heart rates <110 bpm.
Contributors

Astrid N L Hermans
Author
Maastricht University Medical Centre (MUMC) Maastricht , Netherlands (The)

Jonas L Isaksen*
Author

Monika Gawalko
Author

Nikki A H A Pluymaekers
Author
Maastricht University Medical Centre (MUMC) Maastricht , Netherlands (The)

Rachel M J van der Velden
Author

Hilco Snippe
Author

Stijn Evens
Author

Glenn De Witte
Author

Justin G L M Luermans
Author

Martin Manninger
Author

Joost Lumens
Author
Cardiovascular Research Institute Maastricht (CARIM) Maastricht , Netherlands (The)

Jørgen K Kanters
Author

Dominik Linz
Author
You may be interested in