Mapping-guided atrial lead placement determines optimal conduction across Bachmann’s bundle: a rationale for patient-tailored pacing therapy
EP Europace Journal

Abstract
Conventional right atrial appendage (RAA) pacing is associated with increased atrial activation time resulting in higher incidences of atrial tachyarrhythmia. Optimal pacing sites ideally shorten inter-atrial conduction delay, thereby decreasing atrial excitation time. We therefore examined the impact of programmed electrical stimulation (PES) from the right atrium (RA) and left atrium (LA) on the electrophysiological properties of Bachmann’s bundle (BB).
High-resolution epicardial mapping of BB was performed during sinus rhythm (SR) and PES in 34 patients undergoing cardiac surgery. Programmed electrical stimulation was performed from the RAA, junction of the RA with inferior caval vein (LRA), and left atrial appendage (LAA). Pacing from either the RAA or LAA resulted in, respectively, right- and left-sided conduction across BB. However, during LRA pacing in most patients (
Pacing from the LRA results in a remarkable decrease of TAT compared with pacing from the LAA or RAA. As the most optimal pacing site varies between patients, individualized positioning of the atrial pacing lead guided by mapping of BB may be one of the new frontiers for atrial pacing.
Contributors

Nawin L Ramdat Misier
Author

Paul Knops
Author

Annejet Heida
Author

Yannick J H J Taverne
Author

Natasja M S de Groot
Author
You may be interested in