Indication of the radiofrequency induced lesion size by pre-ablation measurements
EP Europace Journal

Abstract
During radiofrequency ablation of arrhythmias tissue heating and hence lesion size depend on electrode–tissue contact and cooling of the electrode tip caused by cavitary blood flow. These factors are unique and unknown for each catheter placement in the beating heart. A tool for assessing these factors prior to ablation may indicate the lesion size which will be obtained for any given catheter position.
Radiofrequency ablation was performed in vitro on strips of left ventricular porcine myocardium during two different levels of convective cooling (0 or 0.1 m/s), two different contact pressures (10 or 30 g) and parallel or perpendicular electrode–tissue orientation using 7F 4 mm tip catheters. Prior to ablation the impedance rise (ΔIMP) caused by the obtained contact and the temperature rise with a 0.6 W 5 s test pulse (Δ
ΔIMP increased significantly (
Pre-ablation measurement of ΔIMP and Δ