Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable defibrillator

EP Europace Journal

5 January 2006
Organised by: Logo
ESC Journals

Abstract

Abstract

This report describes a patient with an implantable defibrillator who suffered an inappropriate defibrillation shock upon retrieving some food items from his inadequately earthed refrigerator. Noise typical of electrical interference can be observed in the stored electrogram of the episode. The patient was instructed to earth his home appliances, but he decided to avoid his refrigerator altogether, and has had no subsequent shocks.