Permanent His-bundle pacing: seeking physiological ventricular pacing

EP Europace Journal

1 April 2010
Organised by: Logo
ESC Journals

Abstract

AbstractAims

Right ventricular apical pacing can have deleterious effects and the His bundle has been widely reported to be an alternative site. This paper presents our experience with permanent His-bundle pacing (HBP).

Methods and results

Patients referred for pacemaker implants (regardless of block type) were screened to determine if temporary HBP corrected conduction dysfunctions (threshold ≤2.5 V for 1 ms) and provided infra-Hisian 1:1 conduction of at least 120 s/m. Of the 182 patients selected, HBP corrected conduction dysfunctions in 133 (73%) patients, 42 (32%) of whom were rejected for the permanent procedure due to high thresholds. His-bundle lead implantation was attempted in the remaining 91 patients and was successful in 59 (65% of all attempts, 44% of all possible cases).

Conclusion

In some patients, permanent HBP may be an alternative to right ventricular apical pacing.