Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation
EP Europace Journal
27 March 2010
Organised by: 

Abstract
Two patients with the same model of single-coil active fix implantable defibrillator lead presented with evidence of lead malfunction, in one case with an abrupt rise in pacing impedance and threshold and in the other with evidence of over-sensing. In both cases, the chest radiograph showed separation of conductors in the defibrillation lead from the main body of the lead.
Contributors

Martyn W. Richards
Author
Princess of Wales Hospital Bridgend , United Kingdom of Great Britain & Northern Ireland

Claire E. Warren
Author

Mark H. Anderson
Author