Automatic adjustment of stimulation output in resynchronization therapy: impact and effectiveness in clinical practice
EP Europace Journal

Abstract
Algorithms for automatic pacing output adjustment have been implemented in pacemakers and implantable defibrillators (ICD) and recently in cardiac resynchronization therapy defibrillators (CRT-D). We assessed the impact and effectiveness of these automatic features.
We prospectively enrolled patients successfully implanted with the following Medtronic CRT-Ds: Concerto [with automatic left ventricular (LV) output management algorithm], Consulta [automatic management of atrial, right ventricular (RV) and LV voltage], and Sentry (only manual voltage adjustments). Patients with complete device data available for at least 12 months were included in the analysis. We analysed data from 739 patients (360 Sentry, 335 Concerto, 44 Consulta). During the first 6 months, the LV pacing amplitude underwent more frequent adjustments in Concerto (63%,
Algorithms for the automatic management of the pacing output reduced pacing output in comparison with the standard manual management approach, with potential optimization of battery longevity. Moreover, they reduced the need to manually reprogram CRT-Ds, suggesting the possibility to simplify CRT-D management and facilitate remote monitoring.