Histological findings around electrodes in pacemaker and implantable cardioverter-defibrillator patients: comparison of steroid-eluting and non-steroid-eluting electrodes
EP Europace Journal

Abstract
To analyse histological findings surrounding the electrodes in pacemaker/implantable cardioverter-defibrillator (PM/ICD) patients. To compare histology around steroid-eluting and non-steroid ventricular pacing electrodes.
In autopsied PM/ICD patients histopathological findings around the electrodes were determined. Seventy patients were studied, PM(58), ICD(12), mean age 75.1 ± 9.3 years. The implantation−death interval was 4.0 ± 3.3 years. Most causes of death were cardiac (PM 52%, ICD 58%). The majority of atrial electrodes were attached to the endocardium and most ventricular electrodes were found in the myocardium (
Atrial electrodes were attached more superficially to the endocardium while PM and ICD ventricular electrodes were more frequently embedded in the myocardium. The electrodes were covered by a connective tissue sheath as a result of thrombus organization. This process persisted most frequently around ICD ventricular electrodes. Only borderline differences were found between the histological findings around steroid-eluting and non-steroid-eluting PM ventricular electrodes.
Contributors

Petr Dvorak
Author

Pavel Kamaryt
Author

Bronislava Slana
Author

Jolana Lipoldova
Author

Pavel Dvorak
Author