Cardiac resynchronization therapy beyond nominal settings: who needs individual programming of the atrioventricular and interventricular delay?
EP Europace Journal

Abstract
This study aimed to determine the additional acute haemodynamic effect of atrioventricular (AV) and interventricular (VV) delay optimization compared with current nominal cardiac resynchronization therapy (CRT) device settings, and to explore whether clinical characteristics correlate to the effect of optimization.
Fifty CRT patients were prospectively enrolled. The optimal AV and VV delays were guided by relative improvement in maximum rate of left ventricular (LV) pressure rise (%d
Compared with the best of the currently available device nominal AV and VV delays, 23–45% of CRT patients can yield additional acute haemodynamic effect by individual optimization of the delays. A new nominal VV delay of 40 ms LV pre-activation is recommended. Male gender, ischaemic aetiology, and longer PR interval are associated with a larger effect of individual optimization.
Contributors

Mathias Meine
Author

Anton E. Tuinenburg
Author

Barun Maskara
Author

Peter Loh
Author

Pieter A. Doevendans
Author