Real-time guidewire localization using impedance-based electroanatomic mapping: experimental results and clinical validation during cryoballoon ablation of atrial fibrillation
EP Europace Journal

Abstract
Cryoballoon ablation is an emerging therapy for atrial fibrillation (AF). However, the Arctic Front cryoballoon (Medtronic) cannot be localized on current electroanatomic mapping (EAM) systems. We describe a technique to visualize guidewires in an impedance-based EAM system.
A novel technique for real-time guidewire localization in an EAM (Ensite Velocity, St Jude Medical) was prospectively evaluated among patients referred for cryoballoon AF ablation. The guidewire was visualized as an ‘orb’ on the EAM and localization in each of the pulmonary veins (PVs) compared with orthogonal fluoroscopy, contrast venography, and intra-cardiac echocardiography. Application of the technique in 21 consecutive patients [median age 58 (interquartile range 21); 71.4% male; 85.7% paroxysmal AF] demonstrated agreement with respect to guidewire localization in 82 of 82 (100%) PVs. Discrimination of guidewire position in the left atrial appendage from the left PVs was also demonstrated. When compared with 21 consecutive cryoballoon procedures over the same time period in which the technique was not used, fluoroscopy time was reduced [median 53.2 (25.9) vs. 72.3 (47.6) min,
This simple, novel technique provides real-time, accurate guidewire localization to enable guidewire and catheter navigation during cryoballoon AF ablation.