Passive-fixation lead failure rates and long-term patient mortality in subjects implanted with Sprint Fidelis electrodes
EP Europace Journal

Abstract
To evaluate passive-fixation lead failure rates and long-term patient survival in subjects implanted with Sprint Fidelis electrodes.
We identified 748 subjects who received a Sprint Fidelis (
Failure rates of passive-fixation Sprint Fidelis leads are increased and similar to those previously reported for active-fixation Fidelis electrodes. Despite the elevated risk for lead failure and its potential sequelae, the Sprint Fidelis has no obvious impact on long-term mortality.