Incidence and characteristics of phrenic nerve palsy following pulmonary vein isolation with the second-generation as compared with the first-generation cryoballoon in 360 consecutive patients
EP Europace Journal

Abstract
The second-generation cryoballoon (CB2) with increased surface cooling has recently become available. The aim was to investigate the incidence and characteristics of phrenic nerve palsy (PNP) during pulmonary vein isolation (PVI) using the CB2 as compared with the first-generation balloon (CB1).
A total of 360 consecutive patients with atrial fibrillation underwent PVI with the CB1 (106 patients) or the CB2 (254 patients). Right PN function was monitored by continuous stimulation and palpation during septal PV ablation. Persistent PNP (present at discharge) occurred in 2.8 and 1.9% (
The rate of transient/persistent PNP associated with the use of the CB2 was 5.9 and 2.8%, respectively. Time to restitution of PN function was longer using the CB2.