Remotely controlled steerable sheath improves result and procedural parameters of atrial fibrillation ablation with magnetic navigation
EP Europace Journal

Abstract
The magnetic navigation (MN) system may be coupled with a new advancement system that fully controls both the catheter and a robotic deflectable sheath (RSh) or with a fixed-curve sheath and a catheter-only advancement system (CAS). We aimed to compare these approaches for atrial fibrillation (AF) ablation.
Atrial fibrillation ablation patients (45, 23 paroxysmal and 22 persistent) performed with MN–RSh (RSh group) were compared with a control group (37, 18 paroxysmal and19 persistent) performed with MN–CAS (CAS group). Setup duration was measured from the procedure's start to operator transfer to control room. Ablation step duration was defined as the time from the beginning of the first radiofrequency (RF) pulse to the end of the last one and was separately acquired for the left and the right pulmonary vein (PV) pairs. Clinical characteristics, left atrial size, and AF-type distribution were similar between the groups. Setup duration as well as mapping times was also similar. Ablation step duration for the left PVs was similar, but was shorter for the right PVs in RSh group (46 ± 9 vs. 63 ± 12 min,
The use of the RSh for AF ablation with MN is safe and improves outcome. Right PV isolation is faster, RF delivery time and procedure time are reduced.