A randomized prospective mechanistic cardiac magnetic resonance study correlating catheter stability, late gadolinium enhancement and 3 year clinical outcomes in robotically assisted vs. standard catheter ablation
EP Europace Journal

Abstract
To prospectively compare cardiac magnetic resonance late gadolinium enhancement (LGE) findings created by standard vs. robotically assisted catheter ablation lesions and correlate these with clinical outcomes.
Forty paroxysmal atrial fibrillation patients (mean age 54 ± 13.8 years) undergoing first left atrial ablation were randomized to either robotic-assisted navigation (Hansen Sensei® X) or standard navigation. Pre-procedural, acute (24 h post-procedure) and late (beyond 3 months) scans were performed with LGE and T2W imaging sequences and percentage circumferential enhancement around the pulmonary vein (PV) antra were quantified. Baseline pre-procedural enhancements were similar in both groups. On acute imaging, mean % encirclements by LGE and T2W signal were 72% and 80% in the robotic group vs. 60% (
Robotically assisted ablation results in greater LGE around the PV antrum. Effective lesions created through improved catheter stability and contact force during initial treatment may have a role in reducing subsequent re-do procedures.