Cavotricuspid isthmus ablation using a catheter equipped with mini electrodes on the 8 mm tip: a prospective comparison with an 8 mm dumbbell-shaped tip catheter and 8 mm tip cryothermal catheter

EP Europace Journal

10 November 2015
Organised by: Logo
ESC Journals

Abstract

AbstractAims

The mini electrodes (ME) placed on the tip of the ablation electrode provide more precise local signal. We evaluated whether ME catheter was effective for the ablation of cavotricuspid isthmus (CTI)-dependent atrial flutter.

Methods and results

Eighty-five consecutive patients (68 men; 62 ± 10 years) underwent CTI ablation either using a catheter equipped with ME on the 8 mm tip (ME catheter) in 25 patients (Group A), 8 mm dumbbell-shaped (DS) tip catheter (DS catheter) in 30 patients (Group B), or 8 mm tip cryothermal catheter (Cryo catheter) in 30 patients (Group C). In cases of failed isthmus block, the catheter was changed to the other catheter, but patients remained in the original group following intention-to-treat analysis. The endpoint was achieved in all patients after 13 ± 7 applications in Group A, 9 ± 4 applications in Group B, and 5 ± 2 applications in Group C (P < 0.001). The fluoroscopic and procedure times were significantly longer in Group A (9 ± 7 and 28 ± 17 min, P = 0.001, and P = 0.002, respectively) when compared with Groups B (6 ± 4 and 13 ± 6 min) and C (4 ± 3 and 14 ± 7 min). A crossover was performed in 14 (56%) Group A patients, and 3 (10%) Group C patients. The mean power delivered in Group A was significantly lower than in Group B (31.3 ± 9.1 vs. 38.6 ± 7.6 W, P = 0.015).

Conclusion

The ME catheter was found to be less effective than the Cryo catheter and a DS catheter for the CTI ablation.

Contributors

Jin Iwasawa
Jin Iwasawa

Author

Tsuchiura Kyodo Hospital Tsuchiura , Japan