Solvent-facilitated lead disconnection for battery replacement in patients with pacemakers or implantable cardioverter defibrillators
EP Europace Journal

Abstract
Battery exchange in pacemaker (PM) or implantable cardioverter defibrillator (ICD) devices may be occasionally problematic because of difficulties in lead disconnection procedures and risk of injuring the fragile leads. This pilot study compares ethanol and dimethyl sulfoxide (DMSO) as solvents to assist removal of leads from PM or ICD device headers in cases of stuck leads or difficulties in untightening device header screws.
Of the total number (527) of our patients requiring battery replacement due to end-of-life (EOL) warnings, conventional exchange was not possible in 34 (6.5%) due to embedding of the lead within blood-derived material. Of these, 30 (17 with PM, 13 with ICD) consented to the study and were randomly assigned to a primary attempt at lead disconnection by ethanol (
Ethanol has utility as a simple and inexpensive modality for lead disconnection from ICD or PM headers.
Contributors
You may be interested in