Reduction of radiation exposure in cryoballoon ablation procedures: a single-centre study applying intracardiac echocardiography and other radioprotective measures
EP Europace Journal

Abstract
The population treated with cryoballoon (CB) ablation is relatively young, and radiation protection is of major importance. We aimed to demonstrate that radiation exposure can be markedly reduced by intracardiac echocardiography (ICE) and optimized settings of the X-ray system.
We analysed 100 patients undergoing CB pulmonary vein isolation (PVI) for treatment of paroxysmal atrial fibrillation. In 50 consecutive patients (25 male, 59 ± 13 years; Group 1), we used ICE, skipped PV angiography prior to CB inflation, and avoided fluoroscopy whenever possible. Furthermore, we reduced the frame rate, minimized distance between patient and detector, and consequently applied collimation. These patients were compared with 50 similar preceding patients in Group 2 (29 male, 61 ± 12 years). Total fluoroscopy time was reduced from 18 ± 6 min in Group 2 to 12 ± 5 min in Group 1 (
Radiation exposure in CB PVI can be markedly reduced without prolonging procedure times, affecting the outcome or complication rates. Moreover, ICE seems to shorten total freezing time.