Contemporary rates and outcomes of single- vs. dual-coil implantable cardioverter defibrillator lead implantation: data from the Israeli ICD Registry
EP Europace Journal

Abstract
Dual-coil leads were traditionally considered standard of care due to lower defibrillation thresholds (DFT). Higher complication rates during extraction with parallel progression in implantable cardioverter defibrillator (ICD) technology raised questions on dual coil necessity. Prior substudies found no significant outcome difference between dual and single coils, although using higher rates of DFT testing then currently practiced. We evaluated the temporal trends in implantation rates of single- vs. dual-coil leads and determined the associated adverse clinical outcomes, using a contemporary nation-wide ICD registry.
Between July 2010 and March 2015, 6343 consecutive ICD (
Real-life rates of single-coil lead implantation are rising while adding no additional risk. These results of single-coil safety are reassuring and obtained, despite low and contemporary rates of DFT testing.