The slope of the initial temperature drop predicts acute pulmonary vein isolation using the second-generation cryoballoon
EP Europace Journal

Abstract
There is no objective, early indicator of occlusion quality, and efficacy of cryoballoon pulmonary vein isolation. As previous experience suggests that the initial cooling rate correlates with these parameters, we investigated the slope of the initial temperature drop as an objective measure.
A systematic evaluation of 523 cryoapplications in 105 patients using a serial ROC-AUC analysis was performed. We found the slope of a linear regression of the temperature–time function to be a good predictor (PPV 0.9, specificity 0.72, sensitivity 0.71, and ROC-AUC 0.75) of acute isolation. It also correlated with nadir temperatures (
About 25 s after freeze initiation, the temperature–time slope predicts important key characteristics of a cryoablation, such as nadir temperature. The slope is the only reported predictor to actually precede acute isolation and thus to support decisions about pull-down manoeuvres or aborting a cryoablation early on. It is also predictive of very low nadir temperatures and phrenic nerve palsy and thus may add to patient safety.
Contributors

Nikolas Deubner
Author

Harald Greiss
Author

Ersan Akkaya
Author

Sergey Zaltsberg
Author

Andreas Hain
Author

Alexander Berkowitsch
Author

Norbert Güttler
Author

Malte Kuniss
Author

Thomas Neumann
Author
You may be interested in