Safe automatic one-lead electrocardiogram analysis in screening for atrial fibrillation
EP Europace Journal

Abstract
Screening for atrial fibrillation (AF) using intermittent electrocardiogram (ECG) recordings can identify individuals at risk of AF-related morbidity in particular stroke. We aimed to validate the performance of an AF screening algorithm compared with manual ECG analysis by specially trained nurses and physicians (gold standard) in 30 s intermittent one-lead ECG recordings.
The STROKESTOP study is a mass-screening study for AF using intermittent ECG recordings. All individuals in the study without known AF registered a 30-s ECG recording in Lead I two times daily for 2 weeks, and all ECGs were manually interpreted. A computerized algorithm was used to analyse 80 149 ECG recordings in 3209 individuals. The computerized algorithm annotated 87.1% (
Automatic ECG screening using a computerized algorithm safely identifies normal ECGs in Lead I and reduces the need for manual evaluation of individual ECGs with >85% with 100% sensitivity on an individual basis.
Contributors

Emma Svennberg
Author

Martin Stridh
Author

Johan Engdahl
Author

Faris Al-Khalili
Author

Leif Friberg
Author

Viveka Frykman
Author

Mårten Rosenqvist
Author