Inappropriate shocks in single-chamber and subcutaneous implantable cardioverter-defibrillators: a systematic review and meta-analysis
EP Europace Journal

Abstract
Single-chamber (VR-ICD) and subcutaneous (S-ICD) implantable cardioverter-defibrillators are effective to protect patients against sudden death but expose them to higher risk of inappropriate shock (IS). We sought to quantify the annual rate and influencing factors of ISs in VR- and S-ICDs from the literature.
PubMed, Embase, and Cochrane Library were searched for full text articles with IS rates. Poisson distribution estimated proportion of patients with ISs; rates were annualized based on follow-up duration. Random effects meta-analysis accounted for study-to-study variation. Out of 3264 articles, 16 qualified for the meta-analysis. Across studies, 6.4% [95% confidence interval (CI) 5.1–7.9%] of patients received an IS per year. Meta-regression analyses demonstrated that IS rates were lower in more recent studies [rate ratio (RR) per year: 0.93, 95% CI: 0.87–0.98;
A comprehensive review of the literature shows that 6.4% of patients with ICDs experienced their first IS annually. One of the 16 studies was better than predicted with the lowest reported rate (1.9%) and could not be explained by timing of the study or other covariates.