Automatic remote monitoring utilizing daily transmissions: transmission reliability and implantable cardioverter defibrillator battery longevity in the TRUST trial
EP Europace Journal

Abstract
Benefits of automatic remote home monitoring (HM) among implantable cardioverter defibrillator (ICD) patients may require high transmission frequency. However, transmission reliability and effects on battery longevity remain uncertain. We hypothesized that HM would have high transmission success permitting punctual guideline based follow-up, and improve battery longevity. This was tested in the prospective randomized TRUST trial.
Implantable cardioverter defibrillator patients were randomized post-implant 2:1 to HM (
Automatic remote HM demonstrated robust transmission reliability. Daily transmission load may be sustained without reducing battery longevity. Home Monitoring conserves battery longevity and tracks long term device performance.
ClinicalTrials.gov; NCT00336284.
Contributors

Charles J Love
Author

Robert Schweikert
Author

Philip Moll
Author

Justin Michalski
Author

Andrew E Epstein
Author