Abstract
Dual coil (DC) electrodes are preferred to single coil (SC) electrodes because of an assumed higher shock efficacy. However, DC-electrodes may be associated with an increased difficulty and risk of lead extraction. We aimed to compare SC- and DC-electrodes with respect to the first shock efficacy (FSE) after implantable cardioverter defibrillator (ICD) implantation.
One thousand and seventy-seven patients of the NORDIC ICD trial were randomly assigned to first time ICD implantation with or without defibrillation (DF) testing. The electrode configuration was determined before randomization. One thousand and sixty-seven patients eventually received an ICD, 516 (48.4%) with a SC- and 551 (51.6%) with a DC-electrode. DC-electrodes were preferentially selected in older patients, renal failure, atrial fibrillation, dual chamber, Cardiac Resynchronization Therapy (CRT) devices, angiotensin-converting-enzyme (ACE) inhibitors/angiotensin (AT) receptor blockers and without Sotalol. However, the preference of the investigational site was dominant over clinical parameters. The DF energy at the final electrode position was higher in SC-electrodes (adjusted difference +1.15 J;
Dual coil-electrode selection mainly depends on the preference of the investigational site and seems to be preferred in older patients, renal failure, atrial fibrillation, dual chamber, and CRT devices. Patients with DC-electrodes required less intraoperative system reconfigurations. Dual coil-electrodes provided a substantially higher FSE during follow-up. Mortality rates were not significantly different in patients with DC- and SC-electrodes.
Contributors

Hendrik Bonnemeier
Author

Johan Brandt
Author

Frank Bode
Author

Jesper Hastrup Svendsen
Author

Oliver Ritter
Author

Johannes Aring
Author

Klaus-Jürgen Gutleben
Author

Ralph Schneider
Author

Angelika Felk
Author

Tino Hauser
Author

Anika Buchholz
Author

Gerhard Hindricks
Author

Karl Wegscheider
Author