Extended use of the wearable cardioverter-defibrillator in patients at risk for sudden cardiac death
EP Europace Journal

Abstract
Data on outcomes in patients using the wearable cardioverter-defibrillator (WCD) > 90 days are limited. We aimed to analyse the clinical course of patients with WCD use ≤90 days vs. WCD use >90 days.
We assessed arrhythmia events during WCD use, and ejection fraction (EF) improvement/implantable cardioverter-defibrillator (ICD) implantation at the end of WCD use in patients with WCD use ≤90 days vs. WCD use >90 days enrolled in the WEARIT-II registry, further assessed by disease aetiology (ischaemic vs. non-ischaemic vs. congenital/inherited heart disease). There were 981 (49%) patients with WCD use >90 days, and 1019 patients with WCD use ≤90 days (median 120 vs. 55 days). There was a lower incidence of sustained ventricular tachycardia/ventricular fibrillation (VT/VF) events (11 vs. 50 events per 100 patient-years,
In WEARIT-II, patients with extended WCD use >90 days remain at risk for ventricular arrhythmia events. One-third of the patients with WCD use >90 days further improved their EF, avoiding the need to consider ICD implantation.