Subcutaneous implantable cardioverter-defibrillator: is it ready for use in children and young adults? A single-centre study
EP Europace Journal

Abstract
Use of the subcutaneous implantable cardioverter-defibrillator (S-ICD) to prevent sudden cardiac death is increasing. Few data exist on S-ICD in young patients. We reviewed our single-centre experience in order to assess the effectiveness and safety of S-ICD implantation, and to identify potential factors associated with complications.
Observational, non-randomized, standard-of-care study on S-ICD implantation/follow-up in young patients with inherited arrhythmias (IA), cardiomyopathies, and congenital heart defects (CHD). Fifteen patients (6 CHD, 8 cardiomyopathies, and 1 IA), median age 15 years (25th–75th centiles, 14–28), 10 of them <18 years, with body mass index (BMI) 22.6 ± 3.4, underwent S-ICD implantation (primary prevention 93%). The first six patients underwent a standard implantation procedure (three surgical incisions), the following nine a two-incision procedure. No intraoperative complications occurred. Over 12 (7–24) months follow-up, two patients received appropriate shocks and one (7%) inappropriate shock. Four patients had device-related complications requiring surgical intervention: three skin erosions at the superior parasternal incision, one pocket infection. A higher risk of complications was seen in patients who underwent standard procedures [hazard ratio (HR) 4.98, 95% confidence interval (CI) 1.53–47.36;
S-ICD implantation was safe and effective in young patients with low rates of inappropriate shock. However, the risk of device-related complications during follow-up remains substantial in patients with low BMI. To prevent possible complications, a two-incision technique should be preferred.
Contributors

Vincenzo Pazzano
Author

Letizia Verticelli
Author

Irma Battipaglia
Author

Fabio Anselmo Saputo
Author

Sonia Albanese
Author

Mariolina Lovecchio
Author

Sergio Valsecchi
Author

Fabrizio Drago
Author
You may be interested in