Real-time magnetic resonance imaging-guided cryoablation of the pulmonary veins with acute freeze-zone and chronic lesion assessment
EP Europace Journal

Abstract
The goals of this study were to develop a method that combines cryoablation with real-time magnetic resonance imaging (MRI) guidance for pulmonary vein isolation (PVI) and to further quantify the lesion formation by imaging both acute and chronic cryolesions.
Investigational MRI-compatible cryoablation devices were created by modifying cryoballoons and cryocatheters. These devices were used in canines (
This study reports on the development of an MRI-based cryoablation system and shows that with acute cryolesions there is a large area of reversible injury. Real-time MRI provides the ability to visualize the freeze-zone formation during the freeze cycle and for focal lesions reaches a maximum after 120 s suggesting that for maximizing lesion size 120 s might be the lower limit for dosing duration.