Monocenter Investigation Micra® MRI study (MIMICRY): feasibility study of the magnetic resonance imaging compatibility of a leadless pacemaker system
EP Europace Journal

Abstract
As
In this prospective non-randomized interventional trial, we enrolled 15 patients with an MRI conditional Micra® leadless pacemaker system to undergo either a 1.5 T or 3.0 T cardiac MRI scan. Clinical adverse events as well as device parameters such as pacing threshold, sensing, impedance, and battery life were assessed at baseline as well as 1 and 3 months after the scan. Device parameter changes between different time points were tested for statistical significance and compared with pre-set cut-off values. Fourteen patients underwent the cardiac MRI scan according to the protocol as well as the scheduled follow-up visits. One participant was excluded from analysis, as the MRI scan was not possible because of severe claustrophobia. Other clinical events did not occur during the scan and the follow-up period. Device parameters stayed stable and changes during the observational period were statistically not significant (changes vs. baseline: pacing threshold: 0.01 ± 0.05 V,
In our set of patients with a Micra® leadless pacemaker, cardiac magnetic resonance imaging at either 1.5 T or 3.0 T proved feasible and safe with no relevant changes in device parameters within 3 months of follow-up.
Contributors

Clemens Steinwender
Author

Juergen Kammler
Author

Hermann Blessberger
Author

Daniel Kiblboeck
Author

Christian Reiter
Author

Thomas Lambert
Author

Joerg Kellermair
Author

Pierre Schmit
Author

Franz Fellner
Author

Michael Lichtenauer
Author

Alexander Kypta
Author
You may be interested in