Atrioventricular optimization in cardiac resynchronization therapy with quadripolar leads: should we optimize every pacing configuration including multi-point pacing?
EP Europace Journal

Abstract
This study aims to define an atrioventricular (AV) delay optimization method for cardiac resynchronization therapy (CRT) with a quadripolar left ventricular (LV) lead based on intrinsic conduction intervals.
Heart failure patients with a left bundle branch block underwent CRT implantation with a quadripolar LV lead. Invasive LV pressure–volume loops were recorded during four biventricular and three multi-point pacing (MPP) settings, using four patient-specific paced AV delays. Haemodynamic response was defined as change in stroke work (Δ%SW) compared to intrinsic rhythm and was related to the following conduction intervals: right atrial pacing to right ventricular sensing interval (RAp-RVs), Q to LV sensing interval normalized to QRS duration (QLV/QRSd), PR-interval, and P-wave duration. In 44 patients, the largest Δ%SW (104 ± 76%) occurred at a paced AV delay of 128 ± 32 ms, at 47 ± 9% of RAp-RVs. Optimal AV delay of biventricular pacing (126 ± 26 ms) did not differ from MPP (126 ± 21 ms,
The AV delay that provides highest haemodynamic response is similar for various LV pacing configurations and for MPP. An AV delay ∼50% of RAp-RVs creates an acute haemodynamic response close to the maximal patient-specific response.