How to perform electroanatomic mapping-guided cardiac resynchronization therapy using Carto 3 and ESI NavX three-dimensional mapping systems
EP Europace Journal

Abstract
To examine the feasibility and safety of a novel protocol for low fluoroscopy, electroanatomic mapping (EAM)-guided Cardiac resynchronization therapy with a defibrillator (CRT-D) implantation and using both EnSite NavX (St. Jude Medical, St. Paul, MN, USA) and Carto 3 (Biosense Webster, Irvine, CA, USA) mapping systems.
Twenty consecutive patients underwent CRT implantation using either a conventional fluoroscopic approach (CFA) or EAM-guided lead placement with Carto 3 and EnSite NavX mapping systems. We compared fluoroscopy and procedural times, radiopaque contrast dose, change in QRS duration pre- and post-procedure, and complications in all patients. Fluoroscopy time was 86% lower in the EAM group compared to the conventional group [mean 37.2 min (CFA) vs. 5.5 min (EAM),
Electroanatomic mapping-guided lead placement using either Carto or ESI NavX mapping systems is a feasible alternative to conventional fluoroscopic methods for CRT-D implantation utilizing the protocol described in this study.