Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator
EP Europace Journal

Abstract
Inappropriate shocks (IAS) remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD). The aims were to assess and characterize different patterns of IAS.
Two hundred and thirty-nine patients were implanted with an S-ICD between 2010 and 2018 for primary and secondary prevention. Follow-up data of at least 6 months were analysed. During a mean follow-up of 34.9 ± 16.0 months, a total of 73 shocks occurred in 38 patients (6%). Forty-three (59%) shocks were considered appropriate due to ventricular tachycardia/ventricular fibrillation, while 30 (41%) were inappropriate and occurred in 19 patients (8%). Myopotentials/noise was the most frequent cause of inappropriate shocks (
Inappropriate therapies are less frequently observed on the alternate vector. The primary vector seems to be unfavourable with regard to oversensing caused by myopotentials. Inappropriate shocks were associated with an increased rate of rehospitalization but not mortality. These observations have implications for the prevention of inappropriate S-ICD shocks.