P1480
French experience of SICD implantation

EP Europace Journal

18 June 2020
Organised by: Logo
ESC Journals

Abstract

AbstractOnBehalf

GPUR

Introduction

The subcutaneous ICD prevents complications of transvenous leads. Its implantation needs a defibrillation test. Performing this test increases the time of procedures..

Purpose

The aim of our study is to describe the experience of subcutaneous ICD of Assitance Publique des Hopitaux de Paris (AP-HP).

Materials and Methods

In a retrospective cohort we included patients who were involved in subcutaneous ICD treatment at the 5 hospitals of AP-HP from December 2012 to April 2018.Analysis used the Kaplan-Meier method and the Mann-Whitney U test.

Results

162 patients were implanted. 76% of implantations were for secondary prevention before 2017, but only 49% after 2017. 126 (77%) tests were successful at first shock, 14 (8.6%) needed a second shock or more, 9 (5.5%) did not have a test. The shock impedance was significantly higher when the shock was not effective at the first test, 82 ohms CI 95% [68; 112] versus 66 ohms CI 95% [64; 70], p <0.05. 6 patients died (3.7%). Late follow up showed 23 (10%) appropriate shocks and 24 inappropriate shocks (11%) mainly due to T wave oversensing (37.5%).

Conclusions

This is the first french experience of SICD implantation. Success of defibrillation test was lower than expected.

Caracteristics of implantation

General anesthesia (%)162 (100)
Procedure time (min)Mean ± SD (median)Range73± 24 (60)20-165
Cameron Health (%)Generator SQ-RX ® 101012 (7.4)
Boston Scientific (%)Generator EMBLEM ® A209100 (61.7)
Generator EMBLEM ® A21944 (27.2)
VF time (sec)Mean ± SD (median)Range16.5± 3.4(15)10-37
Shock Impedance (ohm)Mean ± SD (median) Range72 ± 12.7 (68) 42-130
Success at 1st shock (%)126 (77.7)
Success at 2nd shock or more (%)14 (8.6)
No test performed (%)12 (7.4)

Implantation caracteristics and defibrillation testing.

Impedance was significantly higher in patients without success at first shock : 82 ohms CI 95% [68 ;112] vs 66 ohms CI 95% [64 ;70], p < 0.05.

Abstract Figure. Survival without inappropriate shock